Fluke 189 user manual 500.00 Ω

Models: 189

1 96
Download 96 pages 32.76 Kb
Page 92
Image 92

Model 187 & 189

Users Manual

Function

Resistance 1

Conductance

Range

500.00Ω

5.0000 kΩ

50.000 kΩ

500.00 kΩ

5.0000 MΩ

5.000 MΩ up to 32.000 MΩ

32.0 MΩ up to 50.0 MΩ

50.0 MΩ up to 100.0 MΩ

100.0MΩ up to 500.0 MΩ

50.00nS

Resolution

0.01Ω

0.0001 kΩ

0.001 kΩ

0.01 kΩ

0.0001 MΩ

0.001 MΩ

0.1 MΩ

0.1 MΩ

0.1MΩ

0.01nS

Accuracy

0.05 % + 10 3

0.05% + 2

0.05% + 2

0.05% + 2

0.15% + 4 2

1.0 % + 4 2

3.0 % + 2 4

3.0% + 2 4

10.0 % + 24

1 % + 10

1.For the 5,000 count mode, divide the number of least significant digits (counts) by 10.

2.For relative humidity greater than 70 %, resistance accuracy is 0.5 % over 1 MΩ and 2.5 % over 10 MΩ.

3.Using relative mode (REL Δ) to zero residual reading.

4.To ensure stated accuracy, switch to conductance mode and verify that the open circuit reading is less than 0.10 nS.

7-8

Page 92
Image 92
Fluke 189 user manual 500.00 Ω