Model 187 & 189

Users Manual

Function

Frequency

Duty Cycle

Pulse Width

Temperature

Range

500.00Hz

5.0000 kHz

50.000 kHz

999.99kHz

10.00% to 90.00 %

499.99ms

999.9ms

200 to +1350 ° C

Resolution

0.01 Hz 1

0.0001 kHz

0.001 kHz

0.01kHz

0.01%

0.01ms

0.1ms

0.1° C

Accuracy

±

(0.0050 % + 1)

 

 

 

 

 

 

 

 

±

((voltage range/input voltage) X 300 counts) 5,6

±

(3 % X (voltage range/input voltage) + 1 count) 5,6

 

 

 

 

±

( 1% of reading + 1 ° C) 2,3

MIN MAX AVG

328 to +2462 ° F

0.1 ° F

Response: 100 ms to 80 %

± ( 1% of reading + 1.8 ° F) 2,3

Specified accuracy ± 12 counts for changes > 200 ms in duration. (± 40 counts in AC for changes > 350 ms and inputs > 25 % of range)

FAST MN MX

250 s 4

 

 

1. Reading will be 0.00 for signals below 0.5 Hz.

Specified accuracy ±100 counts up to 5,000 count (full range) reading. For higher peak readings (to 20,000 counts), specified accuracy ± 2% of reading.

2. Accuracy specification is relative to the user-adjustable temperature offset, and assumes ambient temperature stable to ± 1 ° C.

3.For ambient temperature changes of ± 5 ° C, rated accuracy applies after 1 hour.

4.For repetitive peaks; 2.5 ms for single events. Use DC function settings below 20 Hz. 50 mV range not specified.

5.Frequency greater than 5 Hz, except for VDC, 500 mVDC and 3000 mVDC functions; 0.5 Hz to 1 kHz. Signals centered around trigger levels.

6.Range/input ratios also apply to current functions. 500 counts or 5 % for 10 A ranges.

7-10

Page 94
Image 94
Fluke 189 user manual Min Max Avg