Fluke 271 manual Generating Swept Signals, TRIG/SWEEP OUT signal

Models: 271

1 34
Download 34 pages 58.5 Kb
Page 26
Image 26
Generating Swept Signals

271

Getting Started Manual

instrument's AUX OUT socket in order to see the phase effects more clearly.

There are some limitations on using the phase control at frequencies above 30 kHz - chapter 5 of the Users Manual explains these in detail.

Generating Swept Signals

The instrument has an independently-controlled internal trigger generator which produces a square wave with a period from 20 µs to 200 s (in 20 µs steps), corresponding to a repetition frequency from 0.005 Hz to 50 kHz. In most modes the signal is made available at the rear panel TRIG/SWEEP OUT socket.

In sweep mode, the sweep begins on the negative edge of the signal (the transition from +4 to 0 V; if you include markers in the sweep set-up then each marker produces an additional narrow +1 V pulse on the

TRIG/SWEEP OUT signal.

For this tutorial we will generate a logarithmic sine wave sweep from 261.6 to 1108.7 Hz. This 2-octave range is well within the limits of audibility, so you can listen on headphones if you wish. Set the amplitude to about - 10 dBm or you may be surprised by the intensity.

The description "logarithmic" means that the sweep spends an equal period at each note of the scale (and is a misnomer because the rate of frequency increase with time is in fact exponential).

The set up process is as follows:

Press RECALL, 0, ENTER to reset to the factory defaults. Then press EDIT and SWEEP.

16

Page 26
Image 26
Fluke 271 manual Generating Swept Signals, TRIG/SWEEP OUT signal