This instrument meets the requirements of the EMC Directive 89/336/EEC.

Compliance was demonstrated by meeting the test limits of the following standards:

Emissions

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were:

a)

Radiated:

Class B

b)

Conducted:

Class B

c) Harmonics:EN61000-3-2 (2000) Class A

The instrument is Class A by product category.

Immunity

 

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test meth- ods, limits and performance achieved were:

a) EN61000-4-2

(1995)

Electrostatic Discharge: 4 kV air, 4 kV contact

Performance A.

b) EN61000-4-3

(1997)

Electromagnetic Field: 3 V/m, 80 % AM at 1 kHz

Performance A.

c) EN61000-4-11 (1994) Voltage Interrupt: 1 cycle, 100 %

Performance A.

d)EN61000-4-4 (1995) Fast Transient: 1 kV peak (ac line), 0.5 kV peak

(signal lines and RS232/GPIB ports)

Performance A.