51 & 52 Series II

Users Manual

General

80 PK-1 Thermocouple (supplied with

thermometer)

Weight

Dimensions (without holster)

Battery

280 g (10 oz)

2.8cm × 7.8 cm × 16.2 cm (1.1 in × 3 in × 6.4 in)

3 AA batteries

Type

Temperature Range

Accuracy

Type K, Chromel Alumel, bead style

40 oC to +260 oC (40 oF to +500 oF)

±1.1 oC (± 2.0 oF)

Certification

Safety

CAT I

P, )(TÜV pending)

CSA C22.2 No. 1010.1 1992 EN 61010 Amendments 1, 2

OVERVOLTAGE (Installation) CATEGORY I, Pollution Degree 2 per IEC1010-1*

Electrical

Measurement Range

J-type: 210 oC to +1200 oC (346 oF to + 2192 oF)

K-type: 200 oC to +1372 oC (328 oF to +2501 oF)

T-type: 250 oC to +400 oC

*Refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level. Example include protect electronic circuits.

Display Resolution

(418 oF to +752 oF)

E-type: 150 oC to +1000 oC (238 oF to +1832 oF)

0.1oC / oF / K < 1000o

1.0oC / oF / K 1000o

14

Page 18
Image 18