Impulse 6000D/7000DP Users | Manual Supplement |
Pulse playback
+4 V, 1 ms pulse
Start of defibrillator pulse playback [1]
[1]Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early transitional signals.
Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, set the oscilloscope to trigger on the
On page 26, prior to Controlling the Analyzer Remotely, add the following section:
Setting the ECG Reference Lead
ECG signal amplitude is set by designating either lead I or lead II as the reference lead. Lead II is the default selection. To change the reference lead, do the following:
Press Θ to reveal the main setup menu. Next press the softkey labeled More for additional setup selections. Press the softkey labeled ECG Ref Lead to toggle between lead I and lead II. Once the reference lead is set, press the softkey labeled Back or one of the other functions to store the selection in
4/08 | 9 |