Manual Supplement

Impulse 6000D/7000DP Users

Ventricular Fibrillation

Coarse and fine

Monomorphic Ventricular Tachycardia

120 to 300 (by 5) BPM

Polymorphic Ventricular Tachycardia

5 types

Asystole

Flat line

ECG Waves

 

ECG General

 

Lead configuration

12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs

Lead to lead impedance

1000 Ω (nominal)

Rate accuracy

±1 % of nominal

ECG Amplitudes

 

Reference Lead

Selectable, Lead II (default) or Lead I

Settings

0.05 to 0.45 (by 0.05) mV

0.5to 5.0 (by 0.5) mV

Accuracy (All Performance waves and Normal Sinus R waves)

Lead II

 

 

 

±2 %

 

 

 

 

 

 

All other leads

 

 

 

±5 %

 

 

 

 

 

 

Defib paddles

 

 

 

±5 %

 

 

 

 

 

 

Amplitude of ECG signals relative to amplitude setting (in percent)

 

 

 

 

 

Lead II reference

 

 

 

 

 

 

 

 

 

 

Performance waves and R wave detection

 

 

 

 

 

 

 

 

Lead #

I

II

 

III

V1

V2

V3

V4

V5

V6

 

Ref. Amp.

70 %

100 %

 

30 %

100 %

100 %

100 %

100 %

100 %

100 %

4

3/09

Page 5
Image 5
Fluke 6000D, 7000DP manual ECG General, ECG Amplitudes