Impulse 6000D/7000DP Users |
|
|
|
|
|
|
|
| Manual Supplement | ||||
Normal Sinus waves |
|
|
|
|
|
|
|
|
|
|
| ||
| Lead # |
| I | II | III | V1 |
| V2 | V3 | V4 | V5 | V6 |
|
| Ref. Amp. |
| 70 % | 100 % | 30 % | 24 % |
| 48 % | 100 % | 120 % | 112 % | 80 % |
|
Lead I reference |
|
|
|
|
|
|
|
|
|
|
| ||
Performance waves and R wave detection |
|
|
|
|
|
|
|
| |||||
| Lead # |
| I | II | III | V1 |
| V2 | V3 | V4 | V5 | V6 |
|
| Ref. Amp. |
| 100 % | 150 % | 50 % | 100 % |
| 100 % | 100 % | 100 % | 100 % | 100 % |
|
Normal Sinus waves |
|
|
|
|
|
|
|
|
|
|
| ||
| Lead # |
| I | II | III | V1 |
| V2 | V3 | V4 | V5 | V6 |
|
| Ref. Amp. |
| 100 % | 150 % | 50 % | 24 % |
| 48 % | 100 % | 120 % | 112 % | 80 % |
|
ECG Normal | Sinus |
|
|
|
|
|
|
|
|
|
|
|
|
Rates |
| 10 to 360 (by 1) BPM |
|
|
|
|
|
| |||||
ECG High Level Output (BNC Jack) |
|
|
|
|
|
|
|
|
|
| |||
Amplitude |
|
|
|
|
|
|
|
|
|
|
| ||
Range |
| 0.5 V per mV of reference lead setting |
|
|
|
| |||||||
Accuracy |
| ±5 % |
|
|
|
|
|
|
|
| |||
Output Impedance |
| 50 Ω |
|
|
|
|
|
|
|
| |||
ECG on Defibrillator Input Load |
|
|
|
|
|
|
|
|
|
| |||
Same as the LEAD II amplitude but limited to ±4 mV |
|
|
|
|
|
|
|
| |||||
ECG Performance Waves |
|
|
|
|
|
|
|
|
|
|
| ||
Square wave |
| 2.0 and 0.125 Hz |
|
|
|
|
|
| |||||
Triangular wave |
| 2.0 and 2.5 Hz |
|
|
|
|
|
|
8/08 | 5 |