Manual Supplement

Impulse 6000D/7000DP Users

Accuracy

±0.1 ms

Voltage

 

Range

20 to 5000 V

Accuracy

±(1 % of reading + 2 V)

Current

 

Range

0.4 to 100.0 A

Accuracy

±(1 % of reading + 0.1 A)

Tilt (biphasic and pulsed biphasic)

 

Range

1 % to 99 %

Accuracy

±1 digit

Interphase delay (biphasic and pulsed biphasic)

Range

0.1 ms to 9.9 ms

Accuracy

±0.1 ms

Frequency (pulsed biphasic only)

 

Range

2000 Hz to 8000 Hz

Accuracy

±1 % of reading

Duty cycle (pulse biphasic only)

 

Range

1 % to 99 %

Accuracy

±1 digit

Sample rate

250 kHz (4 μs sample)

Maximum Average Power

12 W, equivalent to 10 defib pulses of 360 J every 5 minutes

2

8/08

Page 3
Image 3