Fluke 7000DP, 6000D Replace pages 31 through 35, with the following, Energy Output Measurement

Models: 6000D 7000DP

1 11
Download 11 pages 31.55 Kb
Page 2
Image 2

Impulse 6000D/7000DP Users

Manual Supplement

Change #1, 46470, 46697, 47714, 48007, 49680

Replace pages 31 through 35, with the following:

Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes

Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic

Note

AC Pulsed Biphasic waveform has not been approved in the United States.

Autoranged Measurement

0.1 to 600 J

Accuracy

 

0.1 to 360 J

±(1 % of reading + 0.1 J)

360 to 600 J

±(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

 

Resistance

50 Ω

Accuracy

±1 %, non-inductive (<2 μH)

Pulse trigger level

20 V

Pulse width

 

Range

1.0 to 50.0 ms

4/08

1

Page 2
Image 2
Fluke 7000DP, 6000D manual Replace pages 31 through 35, with the following, Energy Output Measurement, Change #1