Using the Current Output Functions

Manually Stepping the mA Output

Table 9. mA Stepping Pushbuttons

When the rotary switch is in the OUTPUT [mA position, and the OUTPUT jacks are connected to an appropriate load, the meter produces a steady mA dc output. The meter begins sourcing or simulating 0%. Use the pushbuttons to step the current up and down in 25% steps as shown in Table 9. See Table 10 for mA values at each 25% step.

Select either sourcing or simulating by choosing the SOURCE or SIMULATE output jacks.

If the meter cannot deliver the programmed current because the load resistance is too high or the loop supply voltage is too low, dashes (-----) appear on the numeric

display. When the impedance between the SOURCE jacks is low enough, the meter will resume sourcing.

Pushbutton

X

M

%STEP

%STEP T W

Adjustment

Adjusts up to the next higher 25% step

Adjusts down to the next lower 25% step

Note

The COARSE and FINE adjustment pushbuttons described on the previous page are available when you are manually stepping the mA output.

25

Page 29
Image 29
Fluke 787 user manual Manually Stepping the mA Output, MA Stepping Pushbuttons