1 Before You Start

1.1 Introduction

The Hart Scientific 9011 High-AccuracyDual-Well Calibrator may be used as
a portable instrument or bench top temperature calibrator for calibrating ther-
mocouple and RTDtemperature probes. Calibrations may be done over a range
of –30°C to 670°C (–22°F to 1238°F). Temperaturedisplay and setability reso-
lution is 0.1 degrees.
The instrument features:
Rapid heating and cooling
Hot and cold blocks in a single unit
RS-232 interface capability
Switch hold test
Built in programmable features include:
Temperature scan rate control
Eight set-point memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s digital controller.The con-
troller uses a precision platinum RTDas a sensor that controls the well temper-
ature with a triac driven heater or thermoelectric devices (TED).
The LED front panel continuously shows the current well temperature of both
hot and cold blocks. The temperature may be set, using the control buttons, to
any desired temperature within the instrument's range. Multiple fault protection
devices insure user and instrument safety andprotection.
This dry-well calibrator was designed for portability, low cost, and ease of op-
eration. Through proper use, the instrument will provide continued accurate
calibration of temperature sensors and devices. The user should be familiar
with the safety guidelines and operating procedures of the calibrator as de-
scribed in the instruction manual.