Fluke VT02, VT04A user manual Visual Image Alignment Control

Models: VT04 VT02 VT04A

1 40
Download 40 pages 36.69 Kb
Page 17
Image 17

Visual IR Thermometer

Button Operation

Visual Image Alignment Control

The heat map overlay and the visual camera are positioned vertically in the VT Series. This vertical parallax will change with distance to your object. To correct the parallax for a near or far distance, you have a visual image alignment control. See Figure 4.

To toggle the control between a near object or far object:

1.Push NEAR () for a measurement distance from 15 cm to 23 cm (6 in to 9 in).

2.Push FAR () or a measurement distance further than 23 cm/9 in.

The Near or Far icon shows in the upper left corner of the display.

VT02

VT02

9 in / 23 cm

> 9 in / 23 cm

hak19.eps

Figure 4. Visual Image Alignment

9

Page 17
Image 17
Fluke VT02, VT04A user manual Visual Image Alignment Control