3.2 Cleaning the ADF

Feed and Eject rollers

1)Open the ADF when "P" or "0" is not indicated on the Function Number display. If you open the ADF when "P" or "0" is indicated on the Function Number display, the Feed/Eject rollers will not rotate even if you perform step 2) below.

2)Simultaneously hold down the [Send to] and [Scan] buttons on the operator panel. The Feed/Eject rollers start rotating slowly.

3)Put a soft cloth moistened with cleaning fluid to the surface of the rotating Feed/ Eject rollers lightly so that the surface of the rollers is cleaned.

As a guideline, seven presses of the [Send to] and [Scan] buttons make the Feed/ Eject rollers to rotate one full turn.

 

Function No.

 

display

 

Function

 

Send to

 

Scan

 

Power button

 

LED

 

In step 2), the Feed rollers turn at the same time while

￿￿￿￿￿￿￿

rotating the Eject rollers.

 

When cleaning the Eject rollers, be careful not to touch

 

the Feed rollers and vice versa.

Plastic rollers

Clean the Plastic rollers lightly, preventing scratches on the roller surfaces. Take particular care in cleaning these rollers as black debris on them adversely affects the pickup performance. Be careful not to damage the sponges beside the rollers.

Sheet guides Wipe clean lightly.

Glass

Wipe clean lightly.

ATTENTION

If the glass gets dirty, vertical black streaks may appear in the scanned

 

images.

Ultra sonic sensor

Wipe clean lightly with a dry cloth.

3

CARE DAILY

fi-5530C2 Image Scanner Operator’s Guide

 

83