![](/images/new-backgrounds/1075729/7572975x1.webp)
9 Controller Operation
9 Controller Operation
This section discusses in detail how to operate the bath temperature controller using the front control panel. Using the front panel
In the following discussion a button with the word SET, UP, EXIT or DOWN inside indicates the panel button while the dotted box indicates the display reading. Explanation of the button or display reading are to the right of each button or display value.
9.1Bath Temperature
The digital LED display on the front panel allows direct viewing of the actual bath temperature. This temperature value is what is normally shown on the dis- play. The units, C or F, of the temperature value are displayed at the right. For example,
10.00C | Bath temperature in degrees Celsius |
The temperature display function may be accessed from any other function by pressing the “EXIT” button.
9.2Temperature Set-point
The temperature
Setting the temperature involves two steps: (1) select the
(2) adjust the
To protect the refrigeration system from excessive pressures, it will not operate above 50°C.
9.2.1Programmable Set-points
The controller stores 8
35