GE CTR 80 manual Interface Commands, Serial Operation, Sample Period, Duplex Mode, Linefeed

Models: CTR 80

1 61
Download 61 pages 62.35 Kb
Page 52
Image 52

10 Digital Communication Interface

10.1.2.2Sample Period

The sample period is the next parameter in the menu and prompted with

SPEr”. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface ap- proximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press “SET” to choose to set the sample period. Adjust the period with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.

10.1.2.3Duplex Mode

The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the thermometer via the serial interface are immediately echoed or transmitted back to the device of origin. With half du- plex the commands are executed but not echoed. The default setting is half du- plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.

10.1.2.4Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa- rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac- ter (LF, ASCII 10) after transmission of any carriage-return. The default setting is linefeed off. The mode may be changed using “UP” or “DOWN” and press- ing “SET”.

10.1.3Serial Operation

Once the cable has been attached and the interface set up properly the control- ler will immediately begin transmitting temperature readings at the pro- grammed rate. The serial communications uses 8 data bits, one stop bit, and no parity. The set-point and other commands may be sent via the serial interface to set the temperature set-point and view or program the various parameters. The interface commands are discussed in Section10.2. All commands are ASCII character strings terminated with a carriage-return character (CR, ASCII 13).

10.2Interface Commands

The various commands for accessing the calibrator functions via the digital in- terfaces are listed in this section (see Table 3). These commands are used with the RS-232 serial interface. The commands are terminated with a carriage-re- turn character. The interface makes no distinction between upper and lower case letters, hence either may be used. Commands may be abbreviated to the minimum number of letters which determines a unique command. A command may be used to either set a parameter or display a parameter depending on

49

Page 52
Image 52
GE CTR 80 manual Interface Commands, Serial Operation, Sample Period, Duplex Mode, Linefeed