![](/images/new-backgrounds/1075729/75729103x1.webp)
10 Digital Communication Interface
10.1.2.2Sample Period
The sample period is the next parameter in the menu and prompted with
“SPEr”. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface ap- proximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press “SET” to choose to set the sample period. Adjust the period with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
10.1.2.3Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the thermometer via the serial interface are immediately echoed or transmitted back to the device of origin. With half du- plex the commands are executed but not echoed. The default setting is half du- plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
10.1.2.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa- rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac- ter (LF, ASCII 10) after transmission of any
10.1.3Serial Operation
Once the cable has been attached and the interface set up properly the control- ler will immediately begin transmitting temperature readings at the pro- grammed rate. The serial communications uses 8 data bits, one stop bit, and no parity. The
10.2Interface Commands
The various commands for accessing the calibrator functions via the digital in- terfaces are listed in this section (see Table 3). These commands are used with the
49