GE CTR 80 manual Calibration Procedure

Models: CTR 80

1 61
Download 61 pages 62.35 Kb
Page 56
Image 56

11 Calibration Procedure

Set the set-point to 50°C and allow adequate time for the bath to reach this temperature and stabilize. Adjust the DELTA calibration parameter (see Section 9.11.4) to make the bath temperature as measured with the standard thermometer match the set-point. The approximate ratio between a change in DELTA and a change in temperature at 50°C is about 4.0 to 1. For example, if the bath temperature is high by 0.1°C at 50°C then de- crease DELTA by 0.4.

Set the set-point to –70°C and allow adequate time for the bath to reach this temperature and stabilize. Adjust the BETA calibration parameter (see Section 9.11.5) to make the bath temperature as measured with the standard thermometer match the set-point. The approximate ratio between a change in BETA and a change in temperature at –70°C is about –1.0 to 1. For example, if the bath temperature is high by 0.1°C at –70°C then in- crease BETA by 0.1.

54

Page 56
Image 56
GE CTR 80 manual Calibration Procedure