9 Controller Operation
9.8Controller Configuration
The controller has a number of configuration and operating options and calibra- tion parameters which are programmable via the front panel. These are ac- cessed from the secondary menu after the proportional band function by pressing “SET”. Pressing “SET” again enters the first of three sets of configu- ration parameters: operating parameters, serial interface parameters, and cali- bration parameters. The menus are selected using the “UP” and “DOWN” keys and then pressing “SET”. (See Figure 4 on page 36.)
9.9Operating Parameters
The operating parameters menu is indicated by,
PAr | Operating parameters menu |
The operating parameters menu contains the High Limit, Stir Speed, and Cool- ing parameters.
9.9.1High Limit
The High Limit Parameter adjusts the upper
HL | High Limit parameter |
Press “SET” to enable adjustment of HL.
HL | Flashes “HL” and then displays the setting |
H=100 Current HL setting
Adjust the HL parameter using “UP” or “DOWN”.
H=90 | New HL setting |
Press “SET” to accept the new temperature limit.
9.9.2Low Limit
The Low Limit (LL) Parameter adjusts the lower
42