2)The ADD WATER light is on and there is sufficient water in the bath.

With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch.

3)The FAULT light is on.

Check the water temperature. If the water temperature is above 58 °C, the FAULT light will light and the unit will not be operable.

If this temperature is reached, it may indicate a unit malfunction.

Contact GenesisBPS, LLC for further assistance.

4)Unit does not heat. HEATING light is on but actual water temperature is not increasing.

Remove rear cover and check the heating element for continuity. Replace if necessary. This maintenance should only be performed by a trained technician or GenesisBPS, LLC. Contact it or your Biomedical Department.

5)Unit does not rock

Check that the ADD WATER or FAULT lights are not illuminated.

6)Bath moves to side when rocking or the rocking motion is not smooth.

Ensure that the bath is positioned on the white roller.

The unit must be operated on a smooth and level surface. Check the level of the work surface and make adjustments if necessary.