indicator will light. To correct this situation, fill bath with water as explained above.

Add 2 ml of LABath CLEANTM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prevent the formation of turbidity and odor.

2.3Plug the device into a 115 volt AC outlet. For GENESIS 2002, a dedicated 10 Ampere outlet is needed. For the GENESIS 2006, a dedicated 15 Ampere outlet is needed. The outlet must be properly grounded.

IMPORTANT : DO NOT use sharp instruments (pen, pencil, etc.) on the key pad area - it will cause damage to the instrument.

2.4Turn on the POWER SWITCH on the left panel.

2.5Start the unit by pressing the POWER key located on the control panel.

a.The default SET TEMP is 37 °C. When switching on the unit this temperature will be displayed in the SET TEMP window, as the desired temperature, along with the actual WATER TEMP of the bath. The bag timers will display zeros.

b.If the water temperature is less than the set temperature (37 °C) then heating of the water will begin automatically - the HEAT indicator will light. In addition, the HEATING indicator will light, indicating the heater is drawing power. When the bath water reaches the set temperature, HEATING indicator will go off. When the bath water cools, HEATING indicator will go on again, indicating water is being heated. This cycle continues, ensuring accurate water temperature.