c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 °C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other. The display should now show the new SET TEMP.

When the SET TEMP is same as the WATER TEMP, the bath is ready to be used for thawing.

d.Place the frozen product in slot marked 1 on the Plasma Bag Holder.

It is recommended that frozen plasma be placed into a plasma overwrap thawing bag. This should prevent plasma from getting into the bath, should the bag be cracked or damaged and protect the device from damage. After placing frozen plasma into the plasma thawing bag, eliminate all air from bag, twist top of the bag and loop it through the Plasma Bag Holder so it is secure.

For additional overwrap thawing bags, call GenesisBPS, LLC.

It is important to remove the excess air. Place the weight(s) provided into the plasma bag(s) to ensure that the frozen plasma remains submerged in the water.

e. Press the START/CHANGE key for the selected bag location. The timer will start it’s count down while the unit is thawing the plasma. Default timing is 14 minutes for the GENESIS 2002and 18 minutes for the GENESIS 2006. After this time has elapsed an alarm will sound, signaling that the frozen product should now be thawed.

Depending on volume of plasma, the thickness of the plasma thawing bag and storage temperature of product to be thawed,