thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time.
f. To change the default TIMER :
Press the START/CHANGE key for the selected bag location, then press the numeric key pad for the desired minutes. All keys must be pressed within 5 seconds of each other. Minutes must be between 1 and 99.
g.To start rocking motion of the bath, press the ROCK key. Rock indicator will light. Pressing the ROCK key at any time will stop the rocking motion.
NOTE : If the water spills over the top of the bath when rocking motion begins, stop the bath from rocking and remove a small amount of water.
h. To thaw an additional bag, place it in slot 2, 3, etc. Adjust the timer as required, follow instructions in Section
The product should be thawed when an audible signal is heard and the respective BAG TIMER is at zero. Remove the product from that slot location, check the bag contents to ensure that it is completely thawed. If additional thawing is required, return the bag to the holder and set timer for the desired thaw time.
8