LiteRider PTC_Owner’s Manual

Rev A_121911

VII. OPERATION

Battery Gauge Flashes Rapidly (even with the joystick released) See figure 15.

The control system safety circuits have operated and the control system has been prevented from moving the wheelchair. This indicates a system trip, i.e. the VR2 has detected a problem somewhere in the wheelchair’s electrical system.

Please follow this procedure:

Place the joystick in the center position and turn the control system off and on again to clear the fault. If the fault does not clear, go to the next step.

Switch off the control system.

Disconnect and reconnect all connectors on the wheelchair and the control system to make sure they are mated securely.

Check the condition of the battery.

If you can’t find the problem, go to the “Self-Help Guide” section below.

Switch on the control system again and try to drive the wheelchair. If the safety circuits operate again, switch off and do not try to use the wheelchair. Contact your Golden Technologies representative for service.

Self-Help Guide

If a system trip occurs, you can find out what has happened by counting the number of LEDs on the battery gauge that are flashing. See page 25.

Battery Charge

LED Array

 

 

 

 

 

 

On/Off Button

 

 

 

 

 

Speed Setting

 

 

 

Horn Button

 

 

LED Array

 

 

 

 

Speed Setting

 

 

 

 

 

Speed Setting

 

 

 

SLOWER

 

 

 

FASTER

Figure 15

24

Page 25
Image 25
Golden Technologies GP160 owner manual Self-Help Guide