DigiOx
Finger Pulse Oximeter
Operation Manual
Important: Do not operate the John Bunn DigiOx Finger Pulse Oximeter without first reading and understanding this manual! Save this manual for future use.
Note: The most current version of this manual can be found online at www.grahamfield.com
CONTENTS |
|
INTRODUCTION | 3 |
INTENDED USE | 3 |
CONTRAINDICATIONS | 3 |
IMPORTANT SAFETY GUIDELINES - PLEASE |
|
READ BEFORE USE | 3 |
OVERVIEW | 7 |
APPEARANCE | 7 |
FEATURES | 7 |
NAME AND MODEL | 8 |
BATTERY INSTALLATION | 8 |
OPERATION INSTRUCTIONS | 8 |
ACCESSORIES (INCLUDED) | 11 |
REPAIR, MAINTENANCE, AND DISPOSAL | 11 |
MAINTENANCE | 11 |
CLEANING AND DISINFECTION | 12 |
DISPOSAL | 13 |
TROUBLESHOOTING | 13 |
SYMBOL KEY | 15 |
TECHNICAL SPECIFICATIONS | 16 |
LIMITED WARRANTY | 18 |
MANUAL DE OPERACIÓN (ESPAÑOL) | 21 |
2 |
|
INTRODUCTION
Dear User: Please read and understand all instructions before using the John Bunn DigiOx Finger Pulse Oximeter, JB02008.
Intended use
The John Bunn DigiOx Finger Pulse Oximeter is used to measure oxygen saturation (SpO2) and pulse rate in home or clinical settings. It is designed for adult and pediatric patients.
Contraindications
WARNING: The oximeter is not intended for continuous patient monitoring. DO NOT use the oximeter while patient is undergoing MRI or CT scanning.
Important Safety Guidelines - Please Read Before Use The safety statements presented in this chapter refer to the basic safety information that the operator of the John Bunn DigiOx Finger Pulse Oximeter shall pay attention to and abide by. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specific to the operations.
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious injury.
3
Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
Note: Provides application recommendations or other useful information to ensure that you get the most from your product.
WARNINGS:
Important! Read and understand this manual before using the oximeter. Do not use this device without proper instruction from a Healthcare Professional. If the oximeter is not properly operated, personal injury and/or damage to the oximeter could result.
If components are damaged or missing, contact your dealer immediately. DO NOT use substitute parts.
Inspect the device before use to ensure that there is no visible damage that may affect patient or user’s safety or measurement performance. Thereafter, inspect the device at least once a week.
GF Health Products, Inc. assumes no responsibility for any damage or injury caused by improper instal- lation or use of this product.
DO NOT open the oximeter housing. It contains no serviceable parts. Service must be performed by trained, authorized personnel only. Otherwise, device failure and health hazard may occur.
4
EXPLOSION HAZARD: Do not use the oximeter in the presence of flammable anesthetics, explosive substances, vapors or liquids.
The oximeter is designed for
rapid measurement. It is not suitable for
The oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. It is intended as a measuring, not a treatment, device.
Magnetic and electrical fields are capable of interfer- ing with the proper performance of the oximeter. For this reason, ensure that all external devices operated in the vicinity of the oximeter comply with the relevant EMC requirements. Mobile phones,
DO NOT clip this device on edema or tender tissue.
The infrared light emitted from the device (it is invis- ible) is harmful to the eyes. DO NOT stare into the light.
Keep this device out of the reach of children.
5
Cautions
To ensure patient safety, use only parts and ac- cessories specified in this manual.
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
OVERVIEW
Appearance
display screen | battery cover |
power/display |
|
button |
|
from any angle. The oximeter features a low voltage indicator, powers off automatically in eight seconds when not in use, and provides approximately 30 hours of use with the use of alkaline batteries.
Name and Model
Name: John Bunn DigiOx Finger Pulse Oximeter
Avoid placing the device on a limb which is wrapped with a cuff for blood pressure measurement or undergo- ing venous infusion.
Ensure that no obstruction (contamination, scar tissue, fingernail polish, acrylic fingernail, etc.) exists at the site where the sensor is placed; such obstruction could
a. Press the Power/Display Button |
once more; the screen will |
rotate 180° and display as |
shown in Figure 6. |
b. Press the Power/Display |
Button once more; the screen |
Figure 6
If the oximeter gets wet, stop operation im- mediately and do not resume operation until it is dry. When it is carried from a cold environment to a warm and/or humid environment, allow it to warm to room temperature before using it.
DO NOT use sharp objects to press the front panel button.
DO NOT sterilize this device. Do not subject this device to irradiation, steam, ethylene oxide, high temperature, or immersion in liquid. Refer to the Repair and Maintenance/Cleaning and Disinfec- tion section of this manual for instructions for cleaning and disinfection.
| rubber |
| cushion |
Figure 1 | Figure 2 |
The DigiOx Finger Pulse Oximeter is shown above.
Features
The John Bunn DigiOx Finger Pulse Oximeter provides a simple way to
Model: JB02008
BATTERY INSTALLATION
1.Refer to Figure 3; insert two AAA batteries, oriented as shown, into the battery compartment.
2.Replace the cover.
Note: Ensure that the batteries are correctly installed. Incorrect installation will prevent the device from operating.
OPERATION INSTRUCTIONS
Notes:
Figure 3
battery installation
affect the signal received by the sensor and result in an incorrect measurement.
Vigorous exercise and electrosurgical device interfer- ence may affect the measuring accuracy.
1.Open the clip as shown in Figure 4.
2.Insert finger, nail up, into
the clip's rubber cushion (ensure that the finger is
correctly positioned), and then gently clip the finger.
3.Press the Power/Display Button to turn the oxim- eter on.
4.The screen will then display the
data as shown in Figure 5. After |
| will rotate 90° and display as |
| shown in Figure 7. |
c. | Press the Power/Display Button |
| once more; the screen will |
| rotate 180° and display as |
| shown in Figure 8. |
d. Press the Power/Display Button | |
| once more; the screen will |
| rotate |
| as a filled |
| along bottom of screen, as |
| shown in Figure 9. |
e. Press the Power/Display Button | |
| once more; the screen will now |
Figure 7
Figure 8
Figure 9
oxygen saturation and pulse rate. When a finger is inserted into the sensor’s rubber cushion, and the Power/Display Button is pressed to turn the oximeter
Clean the test finger with isopropyl alcohol before and after each measurement. Ensure that the finger is dry before being inserted into the oximeter.
the values become stable, the |
| |
user can read the information | Figure 5 | |
on the display screen. | ||
|
show pulse as a hollow wave- |
form bar graph along bottom of |
screen, as shown in Figure 10. |
Figure 10
6
6.To change the oximeter’s display screen bright- ness, press the Power/Display Button for more than one second. There are 10 brightness levels, and the level, from “BR 1” to “BR 10”, will show on the screen during brightness adjustment. The default is level four. Keep the Power/Display Button depressed for as long as it takes to cycle to the desired brightness level.
7.The oximeter will automatically turn off if no finger is sensed for 8 seconds. If the finger is removed and the Power/Display Button is pressed, the oximeter will remain on and change screens, but will display “FINGER OUT”.
ACCESSORIES (INCLUDED)
1One lanyard
2Two AAA alkaline batteries
3One operation manual
REPAIR, MAINTENANCE, AND DISPOSAL
WARNING: DO NOT open the oximeter housing. It contains no serviceable parts. Service must be performed by trained, authorized personnel only. Oth- erwise, device failure and health hazard may occur.
Maintenance
In order to ensure the oximeter's longevity and continued dependable operation, please observe the following maintenance recommendations.
11 |
TECHNICAL SPECIFICATIONS |
on, then the SpO2 value and pulse rate value display on the screen. The oximeter features a
7
Change the batteries when the
Clean the surface of the device before using: Wipe the device with a clean, soft cloth saturated with a solution of 70% isopropyl alcohol, then let it air dry or wipe it dry. Repeat as necessary.
Take out the batteries if the oximeter will not be used for a long period of time.
Environment (Note: | Operation: 41°F - 104°F |
Humidity may shorten | (5°C - 40°C), 15% - 80% RH |
the life of the device, |
|
or even damage it) | Storage: |
| |
|
|
Cleaning and Disinfection
1.Do not immerse the device in liquid. Clean the de- vice only by using a
2.Wipe with a clean, soft cloth saturated with clean water and air dry, or dry with a clean, soft cloth.
Caution: Do not sterilize this device. Do not subject this device to irradiation, steam, ethylene oxide, high temperature, or immersion in liquid.
Caution: Do not use the sensor if it is dam- aged.
12 |
Do not shake the finger; keep finger motionless during oximeter use.
8
Disposal
Do not treat this device or its batteries as household waste. Separate collection is required.
TROUBLESHOOTING
Problem | Possible Reason | Solution |
|
|
|
SpO2 and | Finger | Reposition finger |
Pulse Rate | incorrectly | correctly and try again |
display | positioned in |
|
unstable or | device |
|
hard to see |
|
|
| Finger is | Ensure patient remains |
| shaking or | immobile |
| patient is |
|
| moving |
|
|
|
|
| Patient’s SpO2 | Remeasure; if |
| value is too low | unsuccessful, yet device |
| to measure | |
| is functioning correctly, | |
|
| |
|
| contact a medical |
|
| professional immediately |
|
|
|
| Excessive | Reduce ambient light |
| ambient light |
|
|
|
|
13
LIMITED WARRANTY
5.When the screen displays as shown in Figure 5, pressing the Power/Display Button will change the screen's orientation:
9
Troubleshooting continued
Problem | Possible Reason | Solution |
|
|
|
The device | Batteries are | Replace batteries |
does not | low |
|
turn on |
|
|
| Batteries | Reinstall batteries |
| incorrectly | correctly |
| installed |
|
|
|
|
| The device has | Contact your GF Health |
| malfunctioned | Products, Inc. distributor |
|
|
|
The device | The device | Normal |
turns off | powers off |
|
suddenly | automatically |
|
| 8 seconds |
|
| after finger is |
|
| removed |
|
|
|
|
| Batteries are | Replace batteries |
| almost drained |
|
|
|
|
Screen | Device | Contact your GF Health |
displays | malfunction | Products, Inc. distributor |
“Error3”, |
|
|
“Error4”, |
|
|
“Error6” or |
|
|
“Error7” |
|
|
|
|
|
| 14 |
|
f.Press the Power/Display Button once more; the screen will return to the original screen as shown in Figure 5 on previous page.
10
SYMBOL KEY
Symbol | Description |
|
|
| Type BF |
|
|
| Warning: See operation manual |
|
|
| The pulse oxygen saturation |
|
|
| Pulse rate (beats per minute) |
|
|
| Low battery voltage |
|
|
SN | Serial number |
| |
| Not for continuous monitoring |
|
|
15
Display screen | Organic LED Display |
|
|
Power supply | Two 1.5V AAA alkaline batteries |
requirement |
|
|
|
SpO2 Parameter | Measuring range: |
Specifications |
|
| Accuracy: |
| no definition |
|
|
Pulse Rate | Measuring range: |
Parameter |
|
Specifications | Accuracy: |
| 235 bpm, ±2% |
|
|
Dimensions (L x | 2.28" x 1.26" x 1.34" (58 mm x 32 mm |
W x H) | x 34 mm) |
|
|
Environment | Operation: 41°F - 104°F (5°C - 40°C), |
| 15% - 80% RH |
|
|
| Storage: |
| < 93% RH |
|
|
Net weight with | 2 oz. (50g) |
batteries |
|
|
|
TECHNICAL SPECIFICATIONS continued | |
|
|
Classification | The type of protection against electric |
| shock: Internally powered equipment |
|
|
| The degree of protection against |
| electric shock: Type BF applied part |
|
|
| The degree of protection against |
| harmful ingress of liquids: Ordinary |
| equipment without protection against |
| ingress of water |
|
|
| |
| I, Class B |
|
|
GF Health Products, Inc. warrants the John Bunn DigiOx Finger Pulse Oximeter, Model JB02008, for a period of one year against manufacturer's defects. If a product is deemed to be under warranty, GF Health Products, Inc. shall provide, at its option, (1) replace- ment of any defective part or product or (2) a credit of the original selling price made to GF Health Products, Inc.'s initial customer. The warranty does not include any labor charges incurred in replacement part(s) in- stallation or any associated freight or shipping charges to GF Health Products, Inc.
Packaging, warranties, products, and specifications are subject to change without notice.
GF Health Products, Inc. is not responsible for typographical errors.
USA Corporate Headquarters:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360
© 2009 GF Health Products, Inc.
Made in China
telephone:
fax:
www.grahamfield.com
16 | 17 | 18 | 19 |