Serviceable Components

Windband

Isolation

Nozzle

Damper

 

Bypass Air

Plenum

Fan Assembly

Weatherhood

Bypass

Damper

Roof Curb or Structural Support

Routine Maintenance & Operation

Once the unit has been put into operation, a routine maintenance schedule should be set up to accomplish the following:

1.Lubrication of bearings and motor.

2.Bypass air dampers should be checked for freedom of operation and wear.

3.Wheel, housing, bolts and set screws on the entire fan should be checked for tightness.

4.Any dirt accumulation on the wheel or in the housing should be removed to prevent unbalance and possible damage.

5.Inspect fan impeller and housing looking for fatigue, corrosion, or wear.

CAUTION

When operating conditions of the fan are to be changed—speed, pressure, temperature, etc.— consult Greenheck to determine if the unit can operate safely at the new conditions.

Fan Operation

All fans should be run every thirty (30) days, or at least “bumped” every thirty days. It is preferred that each fan is run as this causes all electrical and mechanical components to get up to temperature, displacing any formed condensation, redistributes load on bearings, and redistributes grease in the bearings (motor and shaft bearings).

V-Belt Drives

V-belt drives must be checked on a regular basis for wear, tension, alignment and dirt accumulation. Premature or frequent belt failures can be caused by improper belt tension (either too loose or too tight) or misaligned sheaves. Abnormally high belt tension or drive misalignment will cause excessive bearing loads and may result in failure of the fan and/or motor bearings. Conversely, loose belts will cause squealing on start-up, excessive belt flutter, slippage and overheated sheaves. Either excessively loose or tight belts may cause fan vibration.

When replacing V-belts on multiple groove drives all belts should be changed to provide uniform drive loading. Use a set of matched belts whenever possible. Do not pry belts on or off the sheave. Loosen belt tension until belts can be removed by simply lifting the belts off the sheaves. After replacing belts, ensure that slack in each belt is on the same side of the drive. Belt dressing should never be used.

Do not install new belts on worn sheaves. If the sheaves have grooves worn in them, they must be replaced before new belts are installed.

The proper tension for operating a V-belt drive is the lowest tension at which the belts will not slip at peak load conditions. Belts are adjusted by raising or lowering the motor pivot plate. For initial tensioning, the proper belt deflection halfway between sheave centers is 1/64 inch for each inch of belt span.

For example, if the belt span is 64 inches, the belt deflection should be 1 inch using moderate thumb pressure at mid-point of the drive. Check belt tension two times during the first 24 hours of operation and periodically thereafter.

Deflection = Belt Span 64

Belt Span

®

Vektor-CD Centrifugal Laboratory Exhaust 11

Page 11
Image 11
Greenheck Fan 47555 specifications Serviceable Components, Routine Maintenance & Operation, Fan Operation, Belt Drives

47555 specifications

The Greenheck Fan model 47555 represents an innovative solution in air movement technology, designed for various applications requiring efficient and reliable ventilation. This powerful fan embodies advanced engineering and cutting-edge design to optimize airflow while maintaining energy efficiency.

One of the main features of the Greenheck 47555 is its robust construction. Built with high-grade materials, it withstands the rigors of commercial and industrial environments, ensuring longevity and durability. The fan is designed for easy installation and maintenance, which minimizes downtime and operational disruptions.

The 47555 utilizes an advanced airfoil blade design, which enhances performance by maximizing airflow and reducing noise levels. The aerodynamic shape of the blades is specifically engineered to move air efficiently, reducing power consumption while providing optimal ventilation. This characteristic aligns with the growing demand for environmentally friendly solutions in today’s market.

Equipped with a direct drive motor, the Greenheck 47555 offers increased reliability and reduced maintenance needs, as there are fewer moving parts that can wear out over time. The motor is designed to provide peak efficiency at various operating conditions, ensuring that energy usage is minimized without compromising performance.

The fan also includes adjustable speed options, allowing users to tailor airflow to specific requirements. This feature enhances control over ventilation and can contribute to energy savings, especially in applications where variable airflow is needed throughout the day.

Incorporating advanced technology, the Greenheck 47555 is capable of integrating with building management systems for remote monitoring and control. This connectivity enables operators to manage airflow and energy use proactively, aligning with modern smart building initiatives.

Additionally, the unit is available in various sizes and configurations, making it adaptable to a wide range of applications, from commercial kitchens to industrial warehouses. Customization options are also offered to meet unique project specifications.

In summary, the Greenheck Fan model 47555 is a powerful, reliable, and efficient choice for those looking to maintain optimal air quality in their spaces. With its advanced features, durable construction, and energy-efficient design, it meets the demands of modern ventilation needs while contributing to sustainability goals.