Setting up the carousel

MICROLAB® Carousel User Guide (June 2003)

Setting up the carousel

The carousel can be place at the right, left, or front of a ML 4000 base instrument. It is integrated with the ML 4000 via a ML SWAP integration kit.

Carousel housing assembly

If the base plate has not been installed, refer to the “Attaching the base plate to the ML 4000” of the

ML SWAP/4000 Series Integrated System User Guide (7157-01).

To install the carousel housing assembly to the base plate:

1.Place the carousel housing assembly on the base plate, opposite the ML SWAP. Position the housing assembly so that the front end is near the edge of the base plate. See figure 4.

Figure 4: Carousel housing on base plate

2.Determine how close the housing assembly must be to the ML SWAP. There are three positions possible represented by three threaded holes in each side of the housing assembly (see figure 5).

Position 1 is closest to the ML SWAP, and Position 3 is the farthest away. If you plan to use the shelves on the carousel, you must use Position 3. If no shelves will be used, you can use any position, but you may wish to use Position 1 to minimize bench space and maximize throughput.