 TROUBLESHOOTING

SYMPTOMPOSSIBLE CAUSE TROUBLESHOOTING

Refer to the following instructions to check and correct any failure before contacting service personnel.

SYMPTOM

 

POSSIBLE CAUSE

Scale does not turn on

1.

Dead Battery

 

2.

Faulty electrical outlet

 

3.

Bad power supply

Questionable weight or

1.

External object

the scale does not zero

 

interfering with the scale

 

2.

The display did not show

 

 

”0.0” before weighing

 

3.

Scale is not placed on a

 

 

level floor

 

4.

Scale is out of calibration

 

5.

Improper Tare

Weighing is performed but

The patient is not sitting still

the display shows “weigh”

 

and “reweigh” every few

 

seconds; the weighing

 

process takes too long and

 

no weight is displayed.

 

The display shows

The load on the scale

“Overload” message

exceeds the capacity

 

(1000 lbs/450 kgs)

The display shows

The batteries are empty

“LOW BAT” message

 

The display shows

There is a problem with one

“Load Cell Error” message

or more load cells or the load

 

cell cable is disconnected.

CORRECTIVE ACTION

1.Replace batteries

2.Use a different outlet

3.Replace adapter

1.Remove interfering object from the scale

2.Ask the patient to get off the scale, zero the scale and begin weighing process again

3.Place the scale on a level floor and begin weighing process again

4.Check weight with known weight value

5.Place the item to be tared on the scale. Press REWEIGH. Once the weight of the item is displayed, press TARE. Place the patient and the tared item back on the scale. Press REWEIGH again.

Ask the patient to be still or you can change to live weight setting

Remove the excess weight and use the scale according to its limits

Replace batteries according to instructions

Check load cell cable connection at the display and platform assembly ports. If the problem is not corrected, refer to qualified service personnel to replace the defective load cell