Health O Meter 2650KL operation manual Calibration Path, Operator Action, Display

Models: 2650KL

1 71
Download 71 pages 36.73 Kb
Page 21
Image 21
 CALIBRATION PATH

Operator ActionDisplay CALIBRATION PATH

The calibration is performed using kg or lb, according to the units of measure used upon entering into the calibration path.

 

Operator Action

Display

1.

Press ON/OFF button and immediately press and

Calibration

 

hold HOLD/RELEASE button for 3-4 seconds

Process

2.

Press ENTER

Enter load weight

 

 

600.0

3.

Using cd keys and/or the keypad, set the calibration

Zero calibration

 

load to at least 200 lbs (for best results and greater

Please clear the scale

 

accuracy, use a 600 lb weight). Press ENTER

 

 

 

 

4.

Please clear the scale platform and press ENTER

Zero calibration

 

 

Please wait

5.

Please wait without touching or shaking the scale

Weight calibration

 

until the zero calibration process is finished

Put: XXX.X

6.

Load the scale platform with the required weight and

Weight calibration

 

press ENTER

Please wait

7.

Please wait without touching or shaking the scale

Calibration

 

until the calibration process is finished

factor: X.XXXXX

8.

Press ENTER

Calibration

 

 

Please clear the scale

9.

Please remove weight from the scale platform

Rebooting

 

 

Please wait

10.

Please wait until the scale resumes normal operation