TROUBLESHOOTING

If any abnormality arises during use, please check the following points

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

Unit does not turn

Batteries have

Replace them with

on when the START/

run down.

two new AA alkaline

STOP

 

batteries.

button is pushed.

 

 

Battery polarities

Re-insert the batteries

 

have been positioned

in the correct

 

incorrectly.

positions.

 

 

 

EE measurement

Cuff has been placed

Wrap the cuff properly

error symbol shown

incorrectly.

so that it is positioned

on display or the

 

correctly.

blood pressure

 

 

Did you talk or move

Measure again.

value is displayed

during measurement?

Keep arm steady

excessively low

 

during measurement.

Shaking of the arm

(or high).

with the cuff on.

 

 

 

 

 

 

E1 error symbol

Air circuit abnormality.

Check cuff connection.

shown on display

Cuff tube may not be

Measure again.

 

plugged into monitor

 

 

correctly.

 

 

 

 

E2 error symbol

Inflation pressure

Turn the unit off, then

shown on display

exceeding 300 mmHg.

measure again.

 

 

 

E3 error symbol

Error determining

Measure again.

shown on display

measurement data.

 

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.

27

BPS-060_B.indd 29-30

SPECIFICATIONS

Power Source:

Four 1.5V DC (AA) batteries

 

 

Measurement

Oscillometric

Method:

 

Measurement Range:

Pressure: 40~280 millimeters mercury (mmHg)

Pulse: 40~199 beats/minute

Accuracy:

Pressure: ±3 mmHg

Pulse: ±5% of reading

 

Pressure Sensor:

Semi-conductor

 

 

Inflation:

Manual Inflation

Deflation:

Automatic Pressure release valve

Memory Capacity:

60 memories

Auto-shut-off:

1 minute after last button operation

Operation

Temperature: 50°F~104°F (10°C ~40°C)

Environment:

Humidity: 15% – 90% R.H.

Storage Environment:

Temperature: -4°F~158°F (-20°C~70°C)

Humidity: Less than 90% R.H.

Weight:

.43 lbs/193 grams (without batteries)

Arm Circumference:

Standard size adult arm cuff:

9”–13” (23 – 33 cm)

Dimensions:

5”(L) x 4.1”(W) x 2.7”(H)

126mm(L) x 105mm(W) x 68mm(H)

 

 

Internally Powered equipment

 

Type BF Equipment

 

IPx0-No special protection against the

 

ingress of water

 

Not suitable for use in presence of flammable

 

anesthetic mixture with air, oxygen or

 

nitrous oxide.

 

Continuous operation with short-time loading.

 

 

Note: These specifications are subject to change without notice.