4/7/06 | 12:18 PM | Page 19 | |||
| TROUBLESHOOTING/ERROR CODES | ||||
|
|
|
| ||
| Condition | Possible Cause | Corrective Action | ||
|
|
|
|
| |
| Consistent low | Temperature |
| Product | |
| temperature | reading is below |
| range is 50˚F (10°C) to 122˚F | |
| readings (“Lo”) | measurement range | (50°C). Contact your physician if | ||
|
| (below 50˚F/10°C). | you have or suspect any medical | ||
|
|
|
| problems. |
|
|
| The probe is not |
| Make sure probe is properly | |
|
| positioned properly | positioned against the opening | ||
|
| in ear. |
| of the ear canal | |
|
| The waterproof probe | Clean the probe with a | ||
|
| is dirty. |
| cotton swab moistened with | |
|
|
|
| alcohol. |
|
|
|
|
|
| |
| Consistent high | Temperature |
| Product Design- | |
| temperature | reading is above |
| Measurement range is 50˚F (10°C) | |
| readings (“Hi”) | measurement range | to 122˚F (50°C). Contact your | ||
|
| (above 122˚F/50°C). | physician if you have or suspect | ||
|
|
|
| any medical |
|
|
|
|
| problems. |
|
|
|
|
|
| |
| Consistent error | Operating |
| Product Design- | |
| messages (“Err”) | temperature |
| Operating temperature is | |
|
| is not within the |
| 60.8˚F (16°C) to 104˚F (40°C). | |
|
| permissible range for |
|
| |
|
| measurement. |
|
|
|
|
|
|
|
| |
| Low battery warning | Battery power |
| Replace battery (see “Inserting | |
|
| is low. |
| and Replacing Batteries”). | |
|
|
|
|
| |
| Low battery warning | Battery power is |
| Replace battery | |
| with “Lo” symbol | too low for the |
| immediately (see “Inserting and | |
|
| thermometer to |
| Replacing Batteries”). | |
|
| function. |
|
|
|
|
|
|
|
|
|
PRODUCT SPECIFICATIONS
Model Number | Model Number | |
|
| |
Power Source | DC 3 Volt (lithium battery CR2032) | |
|
| |
Battery Life | 1,700 Measurements in day mode or | |
700 Measurements in night mode. | ||
| ||
|
| |
Sensor | Thermopile | |
|
| |
Temperature Indication | In 4 digits, every 0.1˚F or ˚C | |
|
| |
| +/- 0.4˚F/0.2°C between 96.8˚F | |
Accuracy | (36.0°C - 39.0°C) | |
+/- 0.5˚F/0.3°C less than 96.8˚F (36.0°C) or | ||
| ||
| greater than 102.2˚F (39.0°C) | |
|
| |
Measurement Range | 50˚F - 122˚F (10˚C - 50˚C) | |
|
| |
Weight | 0.10 lbs (with battery) | |
|
| |
Operating Temperature | 60.8˚F - 104˚F (16˚C - 40˚C) | |
|
| |
Operating Humidity | Relative humidity up to 95% | |
| ||
|
| |
Storage Temperature | ||
|
| |
Storage Humidity | Relative humidity up to 95% | |
| ||
|
| |
Contents | Ear Thermometer & Battery, Storage Case, | |
Instruction Manual | ||
| ||
|
|
NOTE: These specifications are subject to change without notice. This ear thermometer meets requirements established in ASTM Standard