Agilent Technologies E4406A manual Using the Markers, Troubleshooting Hints

Models: E4406A

1 197
Download 197 pages 57.45 Kb
Page 140
Image 140
Manual background

Making Measurements

Making the Power Stat CCDF Measurement

Using the Markers

The Marker front-panel key accesses the menu to configure the markers.

Select - Allows you to activate up to four markers with the corresponding numbers, respectively. The selected number is underlined and its function is defined by pressing the Function key. The default selection is 1.

Normal - Allows you to activate the selected marker to read the power level and probability of the marker position on the selected curve, for example, which is controlled by the RPG knob.

Delta - Allows you to read the differences in the power levels and probabilities between the selected marker and the next.

Function - Allows you to set the selected marker function to Band Power, Noise, or Off. The default setting is Off. The Band Power and Noise functions are not available for this measurement.

Trace - Allows you to place the selected marker on the Measured, Gaussian, or Reference curve. The default setting is Measured.

Off - Allows you to turn off the selected marker.

Shape - Allows you to access the menu to set the selected marker shape to Diamond, Line, Square, or Cross. The default setting is

Diamond.

Marker All Off - Allows you to turn off all of the markers.

Troubleshooting Hints

The power statistics CCDF measurement can contribute in setting the signal power specifications for design criteria for systems, amplifiers, and other components. For example, it can help determine the optimum operating point to adjust each code timing for appropriate peak/average power ratio throughout the wide channel bandwidth of the transmitter for a W-CDMA system.

As this measurement is a new method, there will be some correlations between CCDF curve degradation and digital radio system measurement parameters such as BER, FER, code domain power, and ACPR. Some studies will help set standards for radio design by specifying the maximum allowed CCDF curve degradation for specific systems.