Using System Configuration and Alignment Keys

NOTE

NOTE

Using System Features

Using System Configuration and Alignment Keys

Using System Configuration and Alignment Keys

There are number of different keys that can help you with understanding the current instrument configuration and for hardware alignment.

Some features are intended for use only by service personnel and cannot be accessed without a service password. If after pressing a key a message appears requesting entry of a password, you should understand this to indicate that this feature is meant for service use only. The password is not available unless the user has purchased the service guide.

Restore System Defaults

Press System, More (1 of 3), More (2 of 3) to access the Restore Sys Defaults key. There are some instrument functions that stay set to the last value that you selected even if you press the instrument preset key or turn the power off and on. The Restore Sys Defaults key allows you to preset those functions back to the original factory settings.

Pressing this key will preset functions like the GPIB/IP address, external reference/trigger settings and auto-alignment selection. You will have to reset these items to return to the your previous setup.

Show Errors

Press System, Show Errors to activate a menu with features that enable you to move around in the error history queue, and to clear the error queue.

Front Panel Key

Level 1

Level 2

 

 

 

System

Show Errors

Top Page

 

 

 

 

 

Last Page

 

 

 

 

 

Next Page

 

 

 

 

 

Prev Page

 

 

 

 

 

Clear Error Queue(s)

 

 

 

 

 

Verbose

 

 

 