Agilent Technologies E4406A manual Setting Up Baseband I/Q Inputs, Table

Models: E4406A

1 197
Download 197 pages 57.45 Kb
Page 161
Image 161
Setting Up Baseband I/Q Inputs

Making Measurements

Using Baseband I/Q Inputs (Option B7C)

Setting Up Baseband I/Q Inputs

Option B7C adds two keys that let you adjust the I/Q inputs; the I/Q Setup key and the I/Q Range key. Both keys are located under the Input/Output front panel key in the Input menu.

The I/Q Range key lets you select one of four levels as an upper limit for the signal being applied to the Baseband I/Q inputs. The level may be selected in units of dBm, dBmv, dBuv, Volts, and Watts. The following table shows the four-level selections available for each unit of measure: The default is 1 Volt.

The I/Q Range power levels in Table 4-8are based on an I/Q Input Z of 50 Ω. I/Q Range voltage levels are independent of I/Q Input Z.

Table 4-8

I/Q Range Settings by Displayed Unit of Measure

 

 

 

 

 

 

 

 

Unit of

Highest

 

 

Lowest

 

Measure

Setting

 

 

Setting

 

 

 

 

 

 

 

dBm

13.0 dBm

7.0 dBm

1.0 dBm

−5.1 dBm

 

 

 

 

 

 

 

dBmv

60 dBmv

54 dBmv

48 dBmv

41.9 dBmv

 

 

 

 

 

 

 

dBuv

120.0 dBuv

114.0 dBuv

108 dBuv

101.9 dBuv

 

 

 

 

 

 

 

Volts

1 Volt

500 mV

250 mV

125 mV

 

 

 

 

 

 

 

Watts

20.0 mW

5.0 mW

1.2 mW

310.0 uW

 

 

 

 

 

 

If I/Q Range is set below the default and the error message “Input Overload” is displayed, this value may be adjusted to its maximum. Beyond that point, the signal must be attenuated to preserve the measurement accuracy. Using a lower value than the default can provide an increased dynamic measurement range.

I/Q Setup lets you adjust the following:

I Offset - Use I Offset to input a value in Volts to offset the measured I value. The default value is 0.0000 Volts, the min/max values are ± 2.5600 Volts. The tuning increment depends on the I/Q Range setting as shown in Table 4-9. This value only affects the displayed results, and does not appear as a correcting voltage at the probe.