I and Q Offset Increment vs. I/Q Range

Making Measurements

Using Baseband I/Q Inputs (Option B7C)

Table 4-9

I and Q Offset Increment vs. I/Q Range

 

 

 

 

 

 

I and Q

 

 

I/Q Range

Offset

 

 

 

Increment

 

 

 

 

 

 

1 Volt

2 mV

 

 

 

 

 

 

500 mV

1 mV

 

 

 

 

 

 

250 mV

.5 mV

 

 

 

 

 

 

125 mV

.25 mV

 

 

 

 

 

Q Offset - Use Q Offset to input a value in Volts to offset the

measured Q value. The default value is 0.0000 Volts, the min/max values are ± 2.5600 Volts. The tuning increment depends on the I/Q Range setting as shown in Table 4-9. This value only affects the displayed results, and does not appear as a correcting voltage at the probe.

I/Q Input Z - This key lets you access a menu to select an input impedance for baseband I/Q input signals. The selection of input

impedance is coupled to a connector “balance” configuration. If I/Q Input Z is set to 1 M Ω, the setting for I/Q Z Ref for Input Z = 1Mkey becomes effective. For details, refer to “Selecting Baseband I/Q Input Impedance” on page 164.

I/Q Z Ref for Input Z = 1M Ω - This key is used to select the

1 M Ω input reference Z value in Ohms. This key is effective only when I/Q Input Z is set to a 1 M Ω setting. The default value is 50.0 . The value range is 0 Ω to 10 M Ω, with a tuning increment of 1.0 . For more details, refer to“Selecting Baseband I/Q Input Impedance” on page 164.

162

Chapter 4

Page 162
Image 162
Agilent Technologies E4406A manual Table, I and Q Offset Increment vs. I/Q Range