Manual background

Making Measurements

Making the Channel Power Measurement

Chan Power Span - Allows you to set the frequency span for the channel power measurement. The range is 1.000 kHz to

10.0000 MHz with 1 Hz resolution. This span is used for the current integration bandwidth setting. Since Chan Power Span is coupled to Integ BW in the factory default condition, if you change the integration bandwidth setting, the channel power span setting changes by a proportional amount, 1.626 times the integration bandwidth, until a limit value is reached. However, the channel power span can be individually set.

Advanced - Allows you to access the following menu to modify the channel power measurement parameters:

Sweep Time - Allows you to manually change the sweep time and

also to toggle the sweep time control between Auto and Man (manual). The range is 1.0 ∝s to 50.00 ms with 1 ∝s resolution. If set to Auto, the sweep time derived from the data point setting is shown on this key regardless of the manual entry range.

Data Points - Allows you to select the number of data points and also to toggle the data point control between Auto and Man (manual). The range is 64 to 65536 with the acceptable entry in powers of 2 (for example: 64, 128, 512). If set to Auto, the optimum number of points is determined for the fastest measurement time with acceptable repeatability. The minimum number of points that could be used is determined by the sweep time and the sampling rate. You can increase the length of the measured time record (capture more of the burst) by increasing the number of points, but the measurement will take longer.

Res BW - Shows information on the resolution bandwidth derived from the sweep time. This key is always grayed out.

Trig Source - Allows you to choose a trigger source from Free Run (Immediate), Video (IF Envlp), RF Burst (Wideband), Ext Front, Ext Rear, Frame, I/Q Input, or Line.