Sensors

Making Measurements

Sensors

Sensors

Purpose

The temperature of the RF board will vary over time and can adversely affect the IF signal amplitude. Sensors on the RF board monitor changes in temperature.

Measurement Method

The sensor measurements are generated internally and do not require any user interaction.

Test Setup

Press Mode, Service, Measure, Sensors.

Results

Window One (IF Signal Level)

Shows a stripchart of the IF signal amplitude in dBm.

Window Two (Cal Osc Level)

The Cal Oscillator window is not implemented.

Window Three (RF Temperature)

Shows a stripchart of the RF assembly temperature in Celsius.

Window Four (numeric results)

IF Signal Level: The ADC number for the detected 21.4 MHz IF signal at the input to the AIF. Typical values with the input attenuator set to 0 dB: 20 (no signal applied); 23(- 10 dBm); 27(- 5 dBm), and 38(0 dBm).

Cal Osc Level: Not implemented.

RF Temp: Shows current temperature in Celsius.