Making Measurements

Making the Waveform (Time Domain) Measurement

Scale Coupling - Allows you to toggle the scale coupling function between On and Off. The default setting is On. Upon pressing the Restart front-panel key or Restart softkey under the Meas Control menu, this function automatically determines the scale per division and reference values based on the measurement results.

If the Signal Envelope window is active in the Signal Envelope view, the AMPLITUDE Y Scale key accesses the menu to modify the following parameters:

Scale/Div - Allows you to set the vertical scale by changing an amplitude value per division. The range is 0.10 to 20.00 dB per division with 0.01 dB resolution. The default setting is 10.00 dB per division. However, since Scale Coupling is defaulted to On, this value is automatically determined by the measurement result.

Ref Value - Allows you to set the reference value ranging from

−250.00 to 250.00 dBm. The default setting is 0.00 dBm. However, since Scale Coupling is defaulted to On, this value is automatically determined by the measurement results.

Ref Position - Allows you to set the reference position to either Top, Ctr (center) or Bot (bottom). The default setting is Top.

Scale Coupling - Allows you to toggle the scale coupling function between On and Off. The default setting is On. Upon pressing the Restart front-panel key or Restart softkey under the Meas Control menu, this function automatically determines the scale per division and reference values based on the measurement results.

If the I/Q Waveform window is active in the I/Q Waveform view, the AMPLITUDE Y Scale key accesses the menu to modify the following parameters:

Scale/Div - Allows you to set the vertical scale by changing an amplitude value per division. The range is 1.00 nV to 20.00 V per division. The default setting is 100.0 mV. However, since Scale Coupling is defaulted to On, this value is automatically determined by the measurement result.

Ref Value - Allows you to set the reference value ranging from

−250.00 to 250.00 V. The default setting is 0.00 V. However, since Scale Coupling is defaulted to On, this value is automatically determined by the measurement results.

Ref Position - Allows you to set the reference position to either Top, Ctr (center) or Bot (bottom). The default setting is Ctr.

Scale Coupling - Allows you to toggle the scale coupling function between On and Off. The default setting is On. Upon pressing the Restart front-panel key or Restart softkey under the Meas Control menu, this function automatically determines the scale per division and reference values based on the measurement results.

Chapter 4

119

Page 119
Image 119
Agilent Technologies E4406A manual Making Measurements, Making the Waveform Time Domain Measurement, Chapter