Alignment

Using System Features

Using System Configuration and Alignment Keys

Alignment

Press System, Alignments to access menus which enable you to align the instrument. Press the ESC key to stop any alignment. The table below diagrams the paths through which you will access alignment feature keys. The feature in the cell that is greyed-out requires a password.

Front

Level 1

Level 2

Level 3

Panel Key

 

 

 

 

 

 

 

System

Alignments

Auto Align

On/Alert/Off

 

 

 

 

 

 

Align All Now

 

 

 

 

 

 

 

Align Subsystems

Align RF

 

 

 

 

 

 

 

Align IF

 

 

 

 

 

 

 

Align ADC

 

 

 

 

 

 

 

Align 50 MHz

 

 

 

Reference

 

 

 

 

 

 

 

Align Current IF

 

 

 

Flatness

 

 

 

 

 

 

 

Align Current

 

 

 

Sysgain

 

 

 

 

 

 

Corrections

On/Off

 

 

 

 

 

 

Time Corr

Auto/On/Off

 

 

 

 

 

 

Restore AlignDefaults

 

 

 

 

 

 

 

Visible Align

Off/Low/High

 

 

 

 

Auto Align - Defaults to the alert state. When set to Alert, the auto align feature causes a message to appear 24 hours after the last alignment to alert tell you that the instrument needs to be aligned. This message will also appear if there is a change of ambient temperature greater than ± 3° C. Use the Align All Now key.

When set to the On state an alignment (which slows down measurements) automatically occurs every 5 minutes or when a change in ambient temperature of greater than ± 3° C has occurred.

Align All Now - Immediately activates an alignment on all of the instrument’s standard RF, IF, and ADC alignments. All other operations are stopped and the alignments progress will be visible on the display.

Chapter 2

77

Page 77
Image 77
Agilent Technologies E4406A manual Alignment, Front, Level, Panel Key