Maximum.

Making Measurements

Making the Adjacent Channel Power (ACP) Measurement

Table 4-4

Adjacent Channel Power Measurement Defaults

 

 

 

 

Measurement Parameter

Factory Default Condition

 

 

 

 

Limit Setup:

 

 

Abs Limit

0.00 dBm

 

Fail

Relative

 

Rel Lim (Car)

−45.00 dBc (offset A)

 

Rel Lim (PSD)

−28.87 dB (offset A)

 

Meas Type

Total Pwr Ref

 

Trig Source

Free Run (Immediate)

 

Fast ACP

Off

 

Limit Test

On

 

Ref Chan Adv:

 

 

Sweep Time

546.1 ∝s; Auto

 

Data Points

2048; Auto

 

Res BW

1.620 kHz (grayed out)

 

Num FFT Seg

1; Auto

 

Offset Adv:

 

 

Sweep Time

11.20 ms; Auto

 

Data Points

1024; Auto

 

Res BW

79.0 Hz (grayed out)

 

Num FFT Seg

1; Auto

 

Relative Atten

0.00 dB

 

 

 

Make sure the ACP measurement is selected under the MEASURE menu. The Meas Setup key accesses the menu which allows you to modify the average number and average mode for this measurement. In addition, the following parameters for adjacent channel power measurements can be modified:

Ref Channel - Allows you to access the following parameters for the reference channel settings:

Chan Integ BW - Allows you to specify the channel integration bandwidth in which the carrier power is measured. The range is 1.000 kHz to 20.0000 MHz with the best resolution of 1 Hz.

Avg Type - Choose the averaging type between Pwr Avg (RMS) and

Maximum.

Ref Chan Adv - Allows you to access the menu to change the following advanced parameters for the reference channel:

Sweep Time - Allows you to select Man or Auto (default) mode.

Chapter 4