remains constant.

Using System Features

Using System Configuration and Alignment Keys

Align Subsystem - Allows you to activate one or a group of the standard internal instrument alignments. Perform alignments on circuitry relating to the following internal alignments. Press the ESC key to cause any alignment to stop.

 

Align RF - Activates an alignment on the RF circuitry.

 

Align IF - Activates an alignment on the IF circuitry. The main

 

gain of the Analog IF is used to compensate for prefilter BW gain

 

variations so the gain of the entire IF path before the A/D

 

remains constant.

 

Align ADC - Activates an alignment on the ADC circuitry.

 

Align 50 MHz Reference - Enables a service technician to

 

adjust the amplitude of the internal 50 MHz reference signal for

 

absolute amplitude accuracy. The test equipment and the

 

adjustment procedure required are described in the E4406A

 

service guide.

 

The Align 50 MHz reference feature can only be activated if you have a

NOTE

 

service password. The password is not available unless the user has

 

purchased the service guide.

 

Align Current IF Flatness - Activates an immediate

 

measurement of the current IF flatness, for FFT spectral

 

amplitude compensation. Normally this alignment occurs in the

 

background when the instrument bandwidth is changed. Given

 

the current BW and Gain DAC settings for a measurement,

 

alignment is done using the comb calibration signal. The relative

 

amplitude of combs within the BW are compared to expected

 

amplitudes to generate the effective shape of the current IF path.

 

If Time Corrections are active, this alignment generates complex

 

IF responses (magnitude and phase) rather than just the original

 

scalar (magnitude) response. The complex alignment requires

 

reasonable delay estimation, and improved trigger compensation

 

so the phase response of several averaged measurements can be

 

unwrapped properly.

 

Align Current SysGain - Activates a fine-tuning adjustment of

 

the system gain. This is done by measuring the response of the

 

current system state configuration to the 50 MHz amplitude

 

reference signal. All subsequent measurements are then

 

compensated appropriately for absolute amplitude accuracy.

 

Normally this occurs in the background when the instrument

 

bandwidth is changed.