Making Measurements

Making the Waveform (Time Domain) Measurement

If the I/Q Polar window is active in the I/Q Polar view, the SPAN X Scale or AMPLITUDE Y Scale key accesses the menu to modify the following parameters:

I/Q Scale/Div - Allows you to set the vertical and horizontal scales by changing a value per division. The range is 1.00 nV to 20.00 V per division. The default setting is 100.0 mV.

I or Q Origin - Allows you to set the reference value ranging from −250.00 to 250.00 V. The default setting is 0.00 V.

The Display key is not available for this measurement.

Using the Markers

The Marker front-panel key accesses the menu to configure the markers.

Select 1 2 3 4 - Allows you to activate up to four markers with the corresponding numbers, respectively. The selected number is underlined and its function is defined by pressing the Function key. The default is 1.

Normal - Allows you to activate the selected marker to read the time position and amplitude of the marker on the RF envelope trace. Marker position is controlled by the RPG knob.

Delta - Allows you to read the differences in time positions and amplitudes between the selected marker and the next.

Function Off - Allows you to define the selected marker function to be

Band Power, Noise, or Off. The default is Off. If set to Band Power, you need to select Delta.

Trace Signal Envelope - Allows you to place the selected marker on

Signal Envelope, I/Q Waveform.

Off - Allows you to turn off the selected marker.

Shape Diamond - Allows you to access the menu to define the selected marker shape to be Diamond, Line, Square, or Cross. The default shape is Diamond.

Marker All Off - Allows you to turn off all of the markers.

The front panel Search key performs a peak search when pressed. A marker will automatically be activated at the highest peak.

120