Power Statistics CCDF Measurement Defaults

Making Measurements

Making the Power Stat CCDF Measurement

Changing the Measurement Setup

The next table shows the factory default settings for power statistics CCDF measurements.

Table 4-6

Power Statistics CCDF Measurement Defaults

 

 

 

 

Measurement Parameter

Factory Default Condition

 

 

 

 

Meas Setup:

 

 

Meas BW

5.00000 MHz

 

Counts

10.0000 Mpoints

 

Meas Interval

1.000 ms

 

Trig Source

Free Run (Immediate)

 

Display:

 

 

Ref Trace

Off

 

Gaussian Line

On

 

 

 

Make sure the Power Stat CCDF measurement is selected under the Measure menu. Press the Meas Setup key to access the menu which allows you to modify the trigger source for this measurement as described in “Measurement Setup” on page 90. In addition, the following parameters can be modified.

Meas BW - Allows you to set the measurement bandwidth according to the channel bandwidth. The range is 10.000 kHz to 6.70000 MHz with 0.1 kHz resolution.

Counts - Allows you to set the accumulated number of sampling points for data acquisition. The range is 1.000 kpoints to

2.000 Gpoints with 1 or 10 kpoints resolution. While this key is activated, enter a value from the numeric keypad by terminating with one of the unit keys shown.

Meas Interval - Allows you to specify the time interval over which the measurement is made. The range is 100.0 ∝s to 10.00 ms with 1 ∝s resolution.

138

Chapter 4

Page 138
Image 138