Model 11793A

Microwave Converter

Equipment

Refer to Table 4, Recommended Test Equipment.

35.Repair Cleanliness

Do not handle the A1 Switch Control Assembly ,,oard more than necessary. If dirt or moisture from the hands comes in contact with the circuit side of the A1 board, the circuits may become inoperative. Do not use solder flux remover on the circuit board.

36.Cleaning the Control-Line Switch Contacts

These instructions pertain to switches S 1 (in instruments with the LO amplifier option), and/or S2

(standard and LO amplifier option configuration) located within the RF module. See Figure 9, Switch Contact Location.

TERMINAL 1

 

 

SPRING CONTACT

TERMINAL C

 

\

 

\

SPRING/ CONTACT

TERMINAL 2

SWITCH.

 

 

CONTACT

 

 

CR2

 

 

Figure 9. Switch Contact Location

1.Remove the switch from the RF module. Refer to Paragraphs 39 or 40, RF Module disassembly procedures.

2.Remove the two screws on the switch cover and remove the cover.

3.Saturate the edge of a sheet of durable paper with alcohol.

27