System performance summaries
HP 8753ES
Configuration |
|
Network analyzer | HP 8753ES |
| Standard and Option 006 |
Calibration kit | HP 85032B |
HP 11857D | |
System dynamic range |
|
30 kHz to 50 kHz | 70 dB1 |
50 kHz to 300 kHz | 90 dB1 |
300 kHz to 16 MHz | 100 dB |
16 MHz to 3 GHz | 110 dB |
Option 014 | 108 dB |
3 GHz to 6 GHz | 105 dB |
Option 014 | 103 dB |
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
1.Typical below 300 kHz.
2.Typical performance.
Measurement port characteristics
Frequency range
Corrected | 30 | 300 | 1.3 | 3 | ||
Directivity |
| 50 dB | 50 dB | 47 dB | 40 dB | |
Source match | 49 dB | 42 dB | 36 dB | 31 dB | ||
Load match |
| 50 dB | 50 dB | 47 dB | 40 dB | |
Reflection |
| ±(0.005 dB | ±(0.009 dB | ±(0.019 dB | ±(0.070 dB | |
tracking |
| +0.02 dB/˚C) | +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) | |
Transmission |
| ±(0.014 dB | ±(0.013 dB | ±(0.026 dB | ±(0.065 dB | |
tracking |
| +0.02 dB/˚C) | +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Phase
Phase
5