System performance summaries

HP 8753ES (50-ohm systems)

Type-N test ports

Configuration

 

Network analyzer

HP 8753ES

 

Standard and Option 006

Calibration kit

HP 85032B

Test-port cables

HP 11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Frequency range

Corrected

30 kHz-300 kHz2

300 kHz-1.3 GHz

1.3 GHz-3 GHz

3 GHz-6 GHz

Directivity

 

50 dB

50 dB

47 dB

40 dB

Source match

49 dB

42 dB

36 dB

31 dB

Load match

 

50 dB

50 dB

47 dB

40 dB

Reflection

 

±(0.005 dB

±(0.009 dB

±(0.019 dB

±(0.070 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Transmission

 

±(0.014 dB

±(0.013 dB

±(0.026 dB

±(0.065 dB

tracking

 

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

5