System performance summaries

HP 8753ET (50-ohm systems)Type-N test ports

Configuration

 

 

 

 

Network analyzer

HP 8753ET Standard,

 

 

Option 006 or Option 004

Calibration kit

 

HP 85032B

Test port cable

 

HP part number 8120-4781

System dynamic range

 

 

 

300 kHz to 16 MHz

100 dB

 

 

16 MHz to 3 GHz

110 dB

 

 

3 GHz to 6 GHz

 

105 dB

 

 

Measurement port characteristics

 

 

 

 

 

 

Frequency range

Uncorrected 300 kHz-1.3 GHz 1.3 GHz-3.0 GHz

3.0 GHz-6.0 GHz

Directivity

30 dB

24 dB

19 dB

Source match4

25 dB

20 dB

14 dB

Load match

24 dB

19 dB

16 dB

Reflection tracking

±1.0 dB

±1.0 dB

±2.0 dB

Transmission tracking

±1.5 dB

±1.5 dB

±2.5 dB

Crosstalk

100 dB

100 dB

90 dB

 

Transmission uncertainty: (enhanced-response calibration)

Transmission uncertainty: (response calibration)

Reflection uncertainty (one-port calibration)

Frequency range (continued)

 

Corrected

300 kHz-1.3 GHz

1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz

 

Reflection measurements1

 

 

 

Directivity

 

50 dB

47 dB

40 dB

Source match

 

42 dB

36 dB

31 dB

Load match

 

 

 

 

 

One-port cal

 

24 dB

19 dB

16 dB

 

Enhanced reflection cal

24 dB

19 dB

16 dB

Reflection tracking

 

±(0.009 dB

±(0.019 dB

±(0.07 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

Transmission measurements2

 

 

Source match

 

 

 

 

 

Enhanced-response cal

42 dB

36 dB

31 dB

 

Response-only cal3

 

25 dB

20 dB

14 dB

Transmission tracking

 

 

 

 

 

Enhanced-response cal

±(0.006 dB

±(0.018 dB

±(0.054 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

 

Response-only cal

 

±(0.033 dB

±(0.1 dB

±(0.27 dB

 

 

 

+0.01 dB/˚C)

+0.02 dB/˚C)

+0.03 dB/˚C)

1.

One-port or enhanced-response calibration.

 

2.

Enhanced-response or response-only calibration.

 

3.

2 dB less with Option 004.

 

4.

Option 004 may degrade uncorrected source match as much as 2 dB.

 

9