System performance summaries

HP 8753ES (75-ohm systems)

Type-F test ports

Configuration

 

Network analyzer

HP 8753E Option 075

Calibration kit

HP 85039B

Test-port cables

HP 11857B

System dynamic range

 

30 kHz to 50 kHz

68 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Data is shown for a Type-F female reflection port and a Type-F male transmission port.

Frequency range

Corrected2

30 kHz-300 kHz

300 kHz-1.3 GHz

1.3 GHz-3 GHz

Directivity

45 dB

45 dB

40 dB

Source match

40 dB

40 dB

30 dB

Load match

45 dB

45 dB

40 dB

Reflection

±(0.060 dB

±(0.060 dB

±(0.024 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

Transmission

±(0.033 dB

±(0.019 dB

±(0.057 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

8