Specifications

Resistance

 

 

 

 

 

 

Max

Range

Resolution

Accuracy

 

Test Current

Open Circuit

 

 

 

 

 

 

Voltage

500 Ω

10 mΩ

± (0.06% + 2)

1

< 800 μA

< 5.5 V

5.0 kΩ

100 mΩ

 

 

 

 

 

 

 

50 kΩ

1Ω

± (0.06% + 2)

 

< 80

μA

 

500 kΩ

10 Ω

 

 

< 15

μA

< 2.2 V

5.0 MΩ

100 Ω

± (0.5% + 1)

 

< 1.5

μA

 

 

50 MΩ

1 kΩ

± (1.0% + 2)

 

< 150 nA

 

 

 

 

 

 

 

 

1After zero adjust of input leads. Zero adjust range up to 0.99 Ω.

Response time: 500 Ω to 500 kΩ — < 2 seconds, 5 M Ω to 50 MΩ — < 10 seconds.

Continuity

Measurement Current: 0.8 mA maximum

Open circuit voltage: < 5.5 Vpeak

Displayed resistance: 0 Ω to 499.99 Ω

Input protection: 500 Vrms (sinewave)

Alarm: Tone when input < 100 Ω ± 50 Ω

Resolution: 10 mΩ (<100 mSec response time)

Diode

Measurement current: +1.0 mA nominal @ 0.6 V

Open circuit voltage: < 5.5 Vpeak

Displayed Voltage: 0 V to 4.999 V

Input protection: 500 Vrms (sinewave)

Accuracy: ± (1% + 2)

Resolution: 100 μV

Frequency (AC Coupled)

Frequency Range

Resolution

Accuracy

Input Voltage (rms)

10 Hz to 99.99 Hz

0.01 Hz

 

 

 

90 Hz to 999.0 Hz

0.1 Hz

 

0.45 mV to 500

V

900 Hz to 9.999 Hz

1 Hz

± (0.05% + 1)

 

 

9.00 kHz to 99.99 kHz

10 Hz

 

.7 V to 100 V

 

90 kHz to 200 kHz

100 Hz

 

1.5 V to 100 V

 

 

 

 

 

 

1-19

Page 19
Image 19
HP Video Game 00974-90002 manual Resistance, Continuity, Diode, Frequency AC Coupled