Specifications

Calibration period: six months minimum. Specifications apply at 23°C ± 5°C, < 80% RH

Accuracy = ± (% of reading + number of digits)

Temperature Coefficient = Accuracy 0.1/° C (0° C to 18° C; 28° C to 55° C)

General

Do not expose product to moisture or rain. Do not use product in flammable atmosphere.

Operating Temperature: 0° to 40°C / 80% RH max (no condensation).

Storage Temperature: -25°C to 60°C / 20% to 70°C RH (no condensation).

Display reading rate: Approximately 2 — 4 times/second

Display rate for frequency measurements: Approximately 1 times/second

Battery life: Approximately 120 hours on DCV

DC Voltage

Range

ResolutionAccuracyInput Resistance

500 mV

10 μV

 

> 1000 MΩ

5 V

100 μV

 

11 MΩ (nominal)

50 V

1 mV

± (0.05% + 2)

 

500 V

10 mV

 

10 MΩ (nominal)

1000 V

100 mV

 

 

 

 

 

 

Normal Mode Rejection Ratio: (NMR) > 60 dB @ 50 or 60 Hz

Effective Common Mode Rejection Ratio (CMR) 1 kΩ imbalance: > 120 dB @ 50 or 60 Hz

AC Voltage (RMS responding, calibrated to display rms)

 

 

 

 

Accuracy

 

 

Input

Range

Resolution

20 Hz to50 Hz to10 kHz to30 kHz to50 kHz toImpedance

 

 

50 Hz10 kHz30 kHz50 kHZ100 kHz(nominal)

500 mV

10 μV

 

± (0.7% +30)

± (2% + 50)

Not Specified

11 M Ω

5 V

100 μV

± (1% + 30)

 

 

 

 

< 50 pF

50 V

1 mV

± (0.5% + 30)

± (1% + 40)

± (2% + 70)

± (3% + 300)

10 M Ω

500 V

10 mV

 

 

 

 

 

 

 

 

 

 

< 50 pF

750 V

100 V

± (1% + 30) 20 Hz to 1 kHz

 

Not Specified

 

 

 

 

 

 

 

 

 

 

 

 

1-17