Calibration Table

Step

Function

Range

Input Signal

Adjustment

Tolerance

(limit)

(counts)

 

 

 

 

17

 

 

48.000 V @ 10 kHz

C3 (±20)

±270

18

 

50 V

48.000 V @ 100 kHz

±1740

19

V

 

48.000 V @ 200 Hz

±270

20

 

500 mV

480.00 mV @ 10 kHz

±366

21

 

750 V

750.0 V @ 200 Hz

±105

22

 

500 Ω

Short

zero adjust 1

±1

23

 

480.00 Ω

9 (±5)

±30

 

 

24

 

5 kΩ

4.8000 kΩ

±30

25

Ω

50 kΩ

48.000 kΩ

±30

26

 

500 kΩ

480.00 kΩ

±30

27

 

5 MΩ

4.8000 MΩ

±242

28

 

50 MΩ

48.000 MΩ

±482

29

500 μA

500 μA

Short

±2

30

480.00 μA

±146

 

 

31

50 mA

50 mA

48.000 mA

±146

32

500 mA

40 mA

480.00 mA

±146

33

10 A

10 A

10.000 A

10 (±10)

±72

34

500 μA

500 μA

480.00 μA @ 200 Hz

±500

35

50 mA

50 mA

48.000 mA @ 200 Hz

±500

36

500 mA

500 mA

480.00 mA @ 200 Hz

±500

37

10 A

10 A

10.00 A @ 200 Hz

±120

38

 

500 Ω

0 Ω to 150 Ω

Tone < 100 Ω

39

 

5 V

1.000 V

±102

40

Hz

5 V

9.000 kHz @ 1 Vrms

±5

1 Perform zero adjustment using key.

6-3