Calibration Table

CAUTION

Dangerous voltages are present during the calibration procedure. Calibration should only be performed by qualified service technicians using a non-conductive tool.

Step

Function

Range

Input SignalAdjustmentTolerance
(limit)(counts)

 

 

 

 

1

 

 

Short

±2

2

500 mV

500 mV

480.0 mV

1 (±2)

±26

3

 

 

-480.0 mV

±26

4

 

50 V

48.000 V

2 (±2)

±26

5

 

-48.000 V

±26

 

 

6

V

5 V

4.800 V

3 (±2)

±26

7

 

500 V

480.00 V

RV4 (±2)

±26

8

 

1000 V

1000 V

RV5 (±1)

±7

9*

V

5 V

4.8000 V @ 200 Hz

6 (±10)

±270

10*

0.2500 V @ 200 Hz

7 (±5)

±42

 

 

11

 

500 mV

480.0 mV @ 200 Hz

8 (±10)

±366

12

 

 

480.00 V @ 10 kHz

C1 (±20)

±270

13

 

500 V

480.00 V @ 100 kHz

±1740

14

 

 

480.00 V @ 200 Hz

±270

15

 

5 V

4 .8000 V @ 10 kHz

C2 (±20)

±270

16

 

4 .8000 V @ 100 kHz

±1740

 

 

*Repeat steps 9 and 10.

6-2